CALIBRASION OF THE ENERGY RESPONSE FOR THE SOFT X-RAY DETECTIONS ELEMENTS WITH THE BEIJING- SYNCHROTRON RADIATION FACILITY
Author(s) -
Sun Ke-Xu,
Rongqing Yi,
Jiamin Yang,
Hongbin Wang,
Hongjuan Ma,
Zhenglin Chen,
Tianxuan Huang,
Yanli Cui,
Zheng Zhi-Jian,
Tang Dao-Yuan,
Yongkun Ding,
Wen Shu-Huai,
Jiang Wen-Mian,
Yuxia Zhao,
Mingqi Cui,
Gang Li,
Chuan-Jue Cui,
Tang E-sheng
Publication year - 1997
Publication title -
wuli xuebao
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.199
H-Index - 47
ISSN - 1000-3290
DOI - 10.7498/aps.46.650
Subject(s) - optics , synchrotron radiation , calibration , beamline , photodiode , photon energy , beam (structure) , radiation , physics , materials science , high energy x rays , monochromator , detector , optoelectronics , photon , quantum mechanics , wavelength
Calibration of the energy response for the soft X-ray detection elements using in the ICF diagnostic are reported.The Beijing synchrotron Radiation Facility (3B1 beam line,photon energy250—1000eV,radiation flux 1012photon/s·mm2·mr2·0.1% band-width,beam current 20—80mA) and target chamber of the reflectometer are used.The source intensity is monitored by AXUV-100 photodiode.The calibration of the energy response for the XRD,filter light film,Ⅱa type diamond photoconductive detector,transmission grating and soft X-ray film are given and analysed.