Rupture of Accura balloon during percutaneous transvenous mitral commissurotomy
Author(s) -
Sunil Kumar Pradhan,
PK Sahoo,
SK Mallick,
Jajati K Padhi
Publication year - 2016
Publication title -
case reports in internal medicine
Language(s) - English
Resource type - Journals
eISSN - 2332-7251
pISSN - 2332-7243
DOI - 10.5430/crim.v4n1p1
Subject(s) - medicine , balloon , percutaneous , surgery , stenosis , radiology , complication , commissurotomy , deformity
Percutaneous transvenous mitral commissurotomy (PTMC) is one of the treatment modality in the management of rheumatic mitral stenosis. Reuse of sterilized PTMC balloon catheters has been widely practiced in developing countries to bring down procedural cost. PTMC by reused balloons are prone to deform and rupture during procedure. This can cause fatal complications like embolism of air or balloon material. Herein, we report a case of Accura balloon rupture during PTMC without any fatal complication. Thus, during balloon preparation particularly of reused balloon, it should be examined for any deformity or tear and air should be removed completely to prevent the fatal outcome.