PRECISION DATA ACQUISITION (DAQ) MODULE WITH REMOTE REPROGRAMMING
Author(s) -
Роман Кочан,
Орест Кочан
Publication year - 2014
Publication title -
computing
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.184
H-Index - 11
eISSN - 2312-5381
pISSN - 1727-6209
DOI - 10.47839/ijc.3.3.309
Subject(s) - data acquisition , computer science , linearity , data collection , computer hardware , calibration , voltage , interface (matter) , real time computing , electronic engineering , electrical engineering , mathematics , statistics , engineering , operating system , bubble , maximum bubble pressure method
These papers are designed precision data acquisition unit based on a 24- bit ADC type AD7712. It is designed for use in distributed information-measuring systems and networks. Flow data collection allows remote reprogramming - a replacement program of work through the serial interface during operation ( after the identification of the appropriate command ). Flow data collection to automatically perform the following procedures designed to improve the accuracy of the measurement result of : (i) the zero ; (ii) the calibration of exemplary source voltage; (iii ) correction of temperature coefficient of exemplary source voltage; (iv ) correction of quadratic integral non-linearity error of the ADC. Implementation procedures improve the accuracy of measurement error provides acceptable DC voltage ± 0.01% in working conditions and non-linearity of the transformation of not more than ±0.0003%.