The “Depression and Elevation” Of ST Segment Settle Down As Momentary “Up and Down “ Of Architect of Artefact : A Case Report Of COVID-19 Patient
Author(s) -
Anurodh Dadarwal,
Sudeep Kumar
Publication year - 2021
Publication title -
journal of cardiology research reviews and reports
Language(s) - English
Resource type - Journals
ISSN - 2634-6796
DOI - 10.47363/jcrrr/2021(2)146
Subject(s) - st segment , artifact (error) , depression (economics) , covid-19 , medicine , st elevation , electrocardiography , chest pain , artificial intelligence , computer science , disease , myocardial infarction , infectious disease (medical specialty) , economics , macroeconomics
A young male was admitted with diagnosis of Covid-19 SARS infection and was having fever, cough & non-cardiac chest pain. There was no past history of cardiac symptoms and physical examination was unremarkable. His vitals were stable. His ECG showed ST elevation and Tall but notched T wave in V3 and ST depression in inferior and lateral leads. Carefully looking these manifestations, they seem the artifacts based on their ECG features and history of the patient. ECG was repeated removing all possible sources of technical errors for these ECG manifestations which showed normal ECG confirming the diagnosis of artifactual ECG. Patient was discharged uneventfully. Recognition of ECG artifact and their technical causes is necessary to avoid inappropriate management.