CALIBRATION OF A SYSTEM FOR ON-LINE MONITORING OF ELECTRON ENERGY AND ABSORBED DOSE AT AN INDUSTRIAL ACCELERATOR LU-10
Author(s) -
R. Pomatsalyuk,
V.Yu. Tytov,
Д. В. Титов,
В.Л. Уваров,
V. Shevchenko
Publication year - 2020
Publication title -
problems of atomic science and technology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.216
H-Index - 17
eISSN - 1562-6016
pISSN - 1682-9344
DOI - 10.46813/2020-127-131
Subject(s) - calibration , absorbed dose , linear particle accelerator , range (aeronautics) , nuclear engineering , aperture (computer memory) , stack (abstract data type) , energy (signal processing) , particle accelerator , physics , beamline , cathode ray , electron , beam (structure) , computer science , materials science , irradiation , optics , nuclear physics , engineering , acoustics , quantum mechanics , composite material , programming language
Particle beam energy and absorbed dose are critical parameters of product processing at industrial electron accelerators. For on-line monitoring of those parameters, a method based on measuring of distribution of the charge induced by irradiation in a wide-aperture stack-monitor, positioned behind an irradiated object, has been developed. A brief review of a control system for monitoring of the processing parameters created with the use of the EPICS package as well as the data of its operating experience at an LU-10 Linac of NSC KIPT are presented in the article. The procedure and results of calibration of the measuring channels within the electron energy range 8…10 MeV are described.