Computed Tomography Angiography with a 192-slice Dual-source Computed Tomography System: Improvements in Image Quality and Radiation Dose
Author(s) -
Philip V.M. Linsen,
Adriaan Coenen,
Marisa M. Lubbers,
Marcel L. Dijkshoorn,
Mohamed Ouhlous,
Koen Nieman
Publication year - 2016
Publication title -
journal of clinical imaging science
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.279
H-Index - 18
eISSN - 2156-7514
pISSN - 2156-5597
DOI - 10.4103/2156-7514.192840
Subject(s) - medicine , image quality , nuclear medicine , spiral (railway) , image noise , contrast to noise ratio , radiation dose , tomography , spiral computed tomography , radiology , computed tomography , image (mathematics) , artificial intelligence , mathematics , mathematical analysis , computer science
Purpose: This study aims to compare image quality, radiation dose, and the influence of the heart rate on image quality of high-pitch spiral coronary computed tomography angiography (CCTA) using 128-slice (second generation) dual-source CT (DSCT) and a 192-slice DSCT (third generation) scanner. Materials and Methods: Two consecutive cohorts of fifty patients underwent CCTA by high-pitch spiral scan mode using 128 or 192-slice DSCT. The 192-slice DSCT system has a more powerful roentgen tube (2 × 120 kW) that allows CCTA acquisition at lower tube voltages, wider longitudinal coverage for faster table speed (732 m/s), and the use of iterative reconstruction. Objective image quality was measured as the signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR). Subjective image quality was evaluated using a Likert scale. Results: While the effective dose was lower with 192-slice DSCT (1.2 ± 0.5 vs. 0.6 ± 0.3 mSv; P < 0.001), the SNR (18.9 ± 4.3 vs. 11.0 ± 2.9; P < 0.001) and CNR (23.5 ± 4.8 vs. 14.3 ± 4.1; P < 0.001) were superior to 128-slice DSCT. Although patients scanned with 192-slice DSCT had a faster heart rate (59 ± 7 vs. 56 ± 6; P = 0.045), subjective image quality was scored higher (4.2 ± 0.8 vs. 3.0 ± 0.7; P < 0.001) compared to 128-slice DSCT. Conclusions: High-pitch spiral CCTA by 192-slice DSCT provides better image quality, despite a higher average heart rate, at lower radiation doses compared to 128-slice DSCT

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom