z-logo
open-access-imgOpen Access
How useful are delta checks in the 21st century? A stochastic-dynamic model of specimen mix-up and detection
Author(s) -
Katie Ovens,
Christopher Naugler
Publication year - 2012
Publication title -
journal of pathology informatics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.009
H-Index - 17
ISSN - 2153-3539
DOI - 10.4103/2153-3539.93402
Subject(s) - computer science , data mining , data science
Delta checks use two specimen test results taken in succession in order to detect test result changes greater than expected physiological variation. One of the most common and serious errors detected by delta checks is specimen mix-up errors. The positive and negative predictive values of delta checks for detecting specimen mix-up errors, however, are largely unknown. Materials and Methods: We addressed this question by first constructing a stochastic dynamic model using repeat test values for five analytes from approximately 8000 inpatients in Calgary, Alberta, Canada. The analytes examined were sodium, potassium, chloride, bicarbonate, and creatinine. The model simulated specimen mix-up errors by randomly switching a set number of pairs of second test results. Sensitivities and specificities were then calculated for each analyte for six combinations of delta check equations and cut-off values from the published literature. Results: Delta check specificities obtained from this model ranged from 50% to 99%; however the sensitivities were generally below 20% with the exception of creatinine for which the best performing delta check had a sensitivity of 82.8%. Within a plausible incidence range of specimen mix-ups the positive predictive values of even the best performing delta check equation and analyte became negligible. Conclusion: This finding casts doubt on the ongoing clinical utility of delta checks in the setting of low rates of specimen mix-ups

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here
Accelerating Research

Address

John Eccles House
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom