Evaluation of PTW Seven29 for tomotherapy patient-specific quality assurance and comparison with ScandiDos Delta4
Author(s) -
Pamela Myers,
Sotirios Stathakis,
Alonso N. Gutiérrez,
Carlos O. Esquivel,
Panayiotis Mavroidis,
Nikos Papanikolaou
Publication year - 2012
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.94741
Subject(s) - tomotherapy , quality assurance , imaging phantom , ionization chamber , nuclear medicine , physics , materials science , medicine , ionization , radiation therapy , radiology , ion , external quality assessment , pathology , quantum mechanics
For routine quality assurance of helical tomotherapy plans, an alternative method, as opposed to the TomoTherapy suggested cylindrical solid water phantom with film and ionization chamber, is proposed using the PTW Seven29 2D-ARRAY inserted in a dedicated octagonal phantom, called Octavius. First, the sensitivity of the array to pitch was studied by varying the pitch during planning to 0.287, 0.433, 1.0, and 2.0. For each pitch selected, the dependence on field size was investigated by generating plans with field widths (FWs) of 1.06 cm, 2.49 cm, and 5.02 cm, for a total of 12 plans. Secondly, a total of 15 patient QA plans were delivered using helical tomotherapy with the Delta(4) and Seven29/Octavius for comparison. Using the clinical gamma criteria, 3% and 3 mm, all FW and pitch plans had a passing percentage of >90%. For patient QA plans, the average gamma pass percentage was 97.0% (94.4-99.8%) for the Delta(4) and 97.6% (92.5-100.0%) for the Seven29/Octavius. Both the Seven29/Octavius and Delta(4) performed to a high standard of measurement accuracy and had a 90% or greater gamma percent for all plans and were considered clinically acceptable.