Performance evaluation of three computed radiography systems using methods recommended in American Association of Physicists in Medicine Report 93
Author(s) -
W.E. Muhogora,
Renato Padovani,
Faustino Bonutti,
P. Msaki,
R. R. Kazema
Publication year - 2011
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.83478
Subject(s) - computed radiography , nuclear medicine , radiography , image quality , medicine , medical physics , optics , radiology , physics , computer science , artificial intelligence , image (mathematics)
The performances of three clinical computed radiography (CR) systems, (Agfa CR 75 (with CRMD 4.0 image plates), Kodak CR 850 (with Kodak GP plates) and Kodak CR 850A (with Kodak GP plates)) were evaluated using six tests recommended in American Association of Physicists in Medicine Report 93. The results indicated variable performances with majority being within acceptable limits. The variations were mainly attributed to differences in detector formulations, plate readers' characteristics, and aging effects. The differences of the mean low contrast scores between the imaging systems for three observers were statistically significant for Agfa and Kodak CR 850A (P=0.009) and for Kodak CR systems (P=0.006) probably because of the differences in ages. However, the differences were not statistically significant between Agfa and Kodak CR 850 (P=0.284) suggesting similar perceived image quality. The study demonstrates the need to implement quality control program regularly.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here