Implementation and validation of a commercial portal dosimetry software for intensity-modulated radiation therapy pre-treatment verification
Author(s) -
Eugenia Moretti,
Varatharaj Chandraraj,
M Ravikumar,
MariaRosa Malisan,
SanjayS Supe,
R. Padovani
Publication year - 2010
Publication title -
journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.71758
Subject(s) - dosimetry , medical physics , radiation therapy , nuclear medicine , software , medicine , computer science , radiology , programming language
Electronic portal imaging devices (EPIDs) are extensively used for obtaining dosimetric information of pre-treatment field verification and in-vivo dosimetry for intensity-modulated radiotherapy (IMRT). In the present study, we have implemented the newly developed portal dosimetry software using independent dose prediction algorithm EPIDose(™) and evaluated this new tool for the pre-treatment IMRT plan quality assurance of Whole Pelvis with Simultaneous Integrated Boost (WP-SIB-IMRT) of prostate cases by comparing with routine two-dimensional (2D) array detector system (MapCHECK(™)). We have investigated 104 split fields using γ -distributions in terms of predefined γ frequency parameters. The mean γ values are found to be 0.42 (SD: 0.06) and 0.44 (SD: 0.06) for the EPIDose and MapCHECK(™), respectively. The average γ∆ for EPIDose and MapCHECK(™) are found as 0.51 (SD: 0.06) and 0.53 (SD: 0.07), respectively. Furthermore, the percentage of points with γ < 1, γ < 1.5, and γ > 2 are 97.4%, 99.3%, and 0.56%, respectively for EPIDose and 96.4%, 99.0% and 0.62% for MapCHECK(™). Based on our results obtained with EPIDose and strong agreement with MapCHECK(™), we may conclude that the EPIDose portal dosimetry system has been successfully implemented and validated with our routine 2D array detector.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom