Performance evaluation of a dedicated computed tomography scanner used for virtual simulation using in-house fabricated CT phantoms
Author(s) -
DS Sharma,
S. D. Sharma,
K.K Sanu,
Sherly Saju,
Deepak D. Deshpande,
S. Kannan
Publication year - 2006
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.25667
Subject(s) - imaging phantom , scanner , materials science , nuclear medicine , biomedical engineering , optics , physics , medicine
Comprehensive tests on single slice CT scanner was carried out using in-house fabricated phantoms/test tools following AAPM recommended methods to independently validate the auto-performance test (APT) results. Test results of all the electromechanical parameters were found within the specified limits. Radiation and sensitivity profile widths were within ± 0.05 cm of the set slice thickness. Effective energy corresponding to nominal kVp of 80, 110 and 130 were 49.99, 55.08 and 59.48 keV, respectively. Percentage noise obtained by APT was 1.32% while the independently measured value was 0.38%. Observed contrast resolutions by independent method at 0.78% and 12% contrast difference were 4 mm and 1.25 mm (= 4 lp/cm) respectively. However, high contrast resolution (limiting spatial resolution) by APT at 50, 10 and 2% MTF levels were 9, 12.5 and 14.1 lp/cm respectively. Difference in calculated and measured CT numbers of water, air, teflon, acrylic, polystyrene and polypropylene were in the range of 0 to 24 HU, while this difference was 46 and 94 HU in case of nylon and bakelite respectively. The contrast scale determined using CT linearity phantom was 1.998×10(-4) cm(-1)/CT number. CT dose index (CTDI) and weighted CTDI (CTDI(w)) measured at different kVp for standard head and body phantoms were smaller than manufacturer-specified and system-calculated values and were found within the manufacturer-specified limit of ± 20%. Measured CTDIs on surface (head: 3.6 cGy and body: 2.6 cGy) and at the center (3.3 cGy, head; and 1.2 cGy, body) were comparable to reported values of other similar CT scanners and were also within the industry-quoted CTDI range. Comprehensive QA and independent validation of APT results are necessary to obtain baseline data for CT virtual simulation.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here