Performance evaluation of a dedicated computed tomography scanner used for virtual simulation using in-house fabricated CT phantoms
Author(s) -
DS Sharma,
S. D. Sharma,
K.K Sanu,
Sherly Saju,
Deepak D. Deshpande,
S. Kannan
Publication year - 2006
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.25667
Subject(s) - imaging phantom , scanner , materials science , nuclear medicine , biomedical engineering , optics , physics , medicine
Comprehensive tests on single slice CT scanner was carried out using in-house fabricated phantoms/test tools following AAPM recommended methods to independently validate the auto-performance test (APT) results. Test results of all the electromechanical parameters were found within the specified limits. Radiation and sensitivity profile widths were within ± 0.05 cm of the set slice thickness. Effective energy corresponding to nominal kVp of 80, 110 and 130 were 49.99, 55.08 and 59.48 keV, respectively. Percentage noise obtained by APT was 1.32% while the independently measured value was 0.38%. Observed contrast resolutions by independent method at 0.78% and 12% contrast difference were 4 mm and 1.25 mm (= 4 lp/cm) respectively. However, high contrast resolution (limiting spatial resolution) by APT at 50, 10 and 2% MTF levels were 9, 12.5 and 14.1 lp/cm respectively. Difference in calculated and measured CT numbers of water, air, teflon, acrylic, polystyrene and polypropylene were in the range of 0 to 24 HU, while this difference was 46 and 94 HU in case of nylon and bakelite respectively. The contrast scale determined using CT linearity phantom was 1.998×10(-4) cm(-1)/CT number. CT dose index (CTDI) and weighted CTDI (CTDI(w)) measured at different kVp for standard head and body phantoms were smaller than manufacturer-specified and system-calculated values and were found within the manufacturer-specified limit of ± 20%. Measured CTDIs on surface (head: 3.6 cGy and body: 2.6 cGy) and at the center (3.3 cGy, head; and 1.2 cGy, body) were comparable to reported values of other similar CT scanners and were also within the industry-quoted CTDI range. Comprehensive QA and independent validation of APT results are necessary to obtain baseline data for CT virtual simulation.