Monte Carlo study of MLC fields for cobalt therapy machine
Author(s) -
Komanduri M. Ayyangar,
Roopa A Rani,
Anil Kumar,
A.R. Reddy
Publication year - 2014
Publication title -
journal of medical physics/journal of medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.292
H-Index - 24
eISSN - 1998-3913
pISSN - 0971-6203
DOI - 10.4103/0971-6203.131279
Subject(s) - collimator , monte carlo method , leakage (economics) , materials science , cobalt 60 , optics , computer science , physics , mathematics , nuclear physics , statistics , irradiation , economics , macroeconomics
An automated Multi-Leaf Collimator (MLC) system has been developed as add-on for the cobalt-60 teletherapy machines available in India. The goal of the present computational study is to validate the MLC design using Monte Carlo (MC) modeling. The study was based on the Kirloskar-supplied Phoenix model machines that closely match the Atomic Energy of Canada Limited (AECL) theratron-80 machine. The MLC is a retrofit attachment to the collimator assembly, with 14 non-divergent leaf pairs of 40 mm thick, 7 mm wide, and 150 mm long tungsten alloy plates with rounded edges and 20 mm tongue and 2 mm groove in each leaf. In the present work, the source and collimator geometry has been investigated in detail to arrive at a model that best represents the measured dosimetric data. The authors have studied in detail the proto-I MLC built for cobalt-60. The MLC field sizes were MC simulated for 2 × 2 cm(2) to 14 × 14 cm(2) square fields as well as irregular fields, and the percent depth dose (PDD) and profile data were compared with ROPS(†) treatment planning system (TPS). In addition, measured profiles using the IMATRIXX system(‡) were also compared with the MC simulations. The proto-I MLC can define radiation fields up to 14 × 14 cm(2) within 3 mm accuracy. The maximum measured leakage through the leaf ends in closed condition was 3.4% and interleaf leakage observed was 7.3%. Good agreement between MC results, ROPS and IMATRIXX results has been observed. The investigation also supports the hypothesis that optical and radiation field coincidence exists for the square fields studied with the MLC. Plots of the percent depth dose (PDD) data and profile data for clinically significant irregular fields have also been presented. The MC model was also investigated to speed up the calculations to allow calculations of clinically relevant conformal beams. (†)Radiation Oncology Planning System (ROPS) is supplied by Tirumala Jyothi Computer Systems described at https://sites.google.com/site/tjcsrops/ (‡)IMATRIXX is supplied by IBA Dosimetry described at HYPERLINK http://www.iba-dosimetry.com.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here