Application of Minimum Effective Cuff Inflating Volume for Laryngeal Mask Airway and its Impact on Postoperative Pharyngeal Complications
Author(s) -
Bingbing Li,
Jie Yan,
Honggang Zhou,
Jing Hao,
Ai-Jia Liu,
Zhengliang Ma
Publication year - 2015
Publication title -
chinese medical journal/chinese medical journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.537
H-Index - 63
eISSN - 2542-5641
pISSN - 0366-6999
DOI - 10.4103/0366-6999.166034
Subject(s) - cuff , medicine , sore throat , laryngeal mask airway , anesthesia , airway , laryngeal masks , surgery , ventilation (architecture) , mechanical engineering , engineering
High intracuff pressure can cause severe pharyngeal complications including sore throat or hoarseness after laryngeal mask airway (LMA) removal postoperatively. Though the application of minimum effective cuff inflating volume is suggested to maintain airway sealing and adequacy of ventilation for patients receiving general anesthesia with LMA at lower level of the intracuff pressure, it is currently not a standard care in most of the anesthetic departments. In this study, the minimum effective cuff inflating volume was determined for classic LMA Well Lead™ (Well Lead Medical Co., Ltd., China) and its impact on postoperative pharyngeal complications was also explored.