Probable edge defect in Acrysof single-piece intraocular lens
Author(s) -
Urmi Shah
Publication year - 2006
Publication title -
indian journal of ophthalmology/indian journal of ophthalmology
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 0.542
H-Index - 51
eISSN - 1998-3689
pISSN - 0301-4738
DOI - 10.4103/0301-4738.27958
Subject(s) - medicine , phacoemulsification , intraocular lens , capsulotomy , forceps , posterior capsulotomy , ophthalmology , lens (geology) , surgery , optics , visual acuity , physics
The purpose of this article is to report an edge defect in single-piece hydrophobic acrylic intraocular lens (IOL) observed during a routine phacoemulsification procedure. The chip was successfully removed intraoperatively with a McPherson's forceps. However, six months postoperatively patient complained of a disturbing glare due to posterior capsular folds originating from the site of IOL defect. Patient was advised, however refused to undergo Nd:Yag laser capsulotomy to relieve the symptoms. This case highlights the necessity to check each foldable IOL under the microscope before loading it into the cartridge, for possible edge defects.