Electron Beam Measurements Employing Electron Montecarlo Algorithm on TrueBeam STx® and Clinac iX® Linear Accelerators
Author(s) -
Kamal Saidi,
Othmane Kaanouch,
Hanae El Gouach,
M.R. Mesradi,
M. Mkimel,
R. El Baydaoui
Publication year - 2021
Publication title -
open access macedonian journal of medical sciences
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.288
H-Index - 17
ISSN - 1857-9655
DOI - 10.3889/oamjms.2021.7564
Subject(s) - truebeam , electron , monte carlo method , physics , linear particle accelerator , cathode ray , medicine , beam (structure) , bremsstrahlung , nuclear medicine , imaging phantom , optics , nuclear physics , statistics , mathematics
Electron beam measurement comparison between TrueBeam STx® and Clinac iX® established. Data evaluation of eMC-calculated and measured for TrueBeam STx® performed. Dosimetric parameters measured including depth dose curves for each applicator, percentage depth dose (PDDs) curves without applicator, the profile in-air for a large field size 40×40 cm2, and the Absolute Dose (cGy/MU) for each applicator using a large water phantom (PTW, Freiburg, Germany), employing Roos and Markus plane-parallel ionization chambers. The data were examined for five electron beams of Varian’s TrueBeam STx® and Clinac iX® machines. A comparison between measurement PDDs and calculated by the Eclipse electron Monte Carlo (eMC) algorithm was performed to validate Truebeam STx® commissioning. The measured data indicated that electron beam PDDs from the TrueBeam STx® machine are well matched to those from Clinac iX® machine. The quality index R50 for applicator 15×15 cm2 was in the tolerance intervals. However, Surface dose (Ds) increases with increasing energy for both accelerators. Comparisons between the measured and eMC-calculated values revealed that the R100, R90, R80, and R50 values mostly agree within 5 mm. Measured and calculated bremsstrahlung tail Rp correlates well statistically. Ds agrees mostly within 2%. Electron beams were successfully validated for TrueBeam STx®, a good agreement between modeled and measured data was observed.