Electron Beam Measurements Employing Electron Montecarlo Algorithm on TrueBeam STx® and Clinac iX® Linear Accelerators
Author(s) -
Kamal Saidi,
Othmane Kaanouch,
Hanae El Gouach,
M.R. Mesradi,
M. Mkimel,
R. El Baydaoui
Publication year - 2021
Publication title -
open access macedonian journal of medical sciences
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.288
H-Index - 17
ISSN - 1857-9655
DOI - 10.3889/oamjms.2021.7564
Subject(s) - truebeam , electron , monte carlo method , physics , linear particle accelerator , cathode ray , medicine , beam (structure) , bremsstrahlung , nuclear medicine , imaging phantom , optics , nuclear physics , statistics , mathematics
Electron beam measurement comparison between TrueBeam STx® and Clinac iX® established. Data evaluation of eMC-calculated and measured for TrueBeam STx® performed. Dosimetric parameters measured including depth dose curves for each applicator, percentage depth dose (PDDs) curves without applicator, the profile in-air for a large field size 40×40 cm2, and the Absolute Dose (cGy/MU) for each applicator using a large water phantom (PTW, Freiburg, Germany), employing Roos and Markus plane-parallel ionization chambers. The data were examined for five electron beams of Varian’s TrueBeam STx® and Clinac iX® machines. A comparison between measurement PDDs and calculated by the Eclipse electron Monte Carlo (eMC) algorithm was performed to validate Truebeam STx® commissioning. The measured data indicated that electron beam PDDs from the TrueBeam STx® machine are well matched to those from Clinac iX® machine. The quality index R50 for applicator 15×15 cm2 was in the tolerance intervals. However, Surface dose (Ds) increases with increasing energy for both accelerators. Comparisons between the measured and eMC-calculated values revealed that the R100, R90, R80, and R50 values mostly agree within 5 mm. Measured and calculated bremsstrahlung tail Rp correlates well statistically. Ds agrees mostly within 2%. Electron beams were successfully validated for TrueBeam STx®, a good agreement between modeled and measured data was observed.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here