Assessment of the fluid dynamic performance of a vitreous cutter
Author(s) -
Alessandro Stocchino,
Rodolfo Repetto,
Mario R. Romano
Publication year - 2018
Publication title -
modeling and artificial intelligence in ophthalmology
Language(s) - English
Resource type - Journals
eISSN - 2468-3930
pISSN - 2468-3922
DOI - 10.35119/maio.v2i2.80
Subject(s) - vitrectomy , tamponade , particle image velocimetry , materials science , ophthalmology , computer science , biomedical engineering , medicine , physics , mechanics , visual acuity , turbulence
Vitreous cutters are surgical devices used during vitrectomy to remove the vitreous humor from the eye and replace it with tamponade fluids. The aim of the present work is to assess the performance of the EVA Phaco-vitrectomy System vitreous cutter (Dutch Ophthalmic Research Center (DORC) BV; Zuidland, The Netherlands) used with diff erent needle sizes and blade shapes. The analysis is based on laboratory measurements of fluid flow performed using the particle image velocimetry (PIV) technique.