Calculation of the shielding safety of Computed Tomography scanner room by using MCNP5 code
Author(s) -
Khanh Ai Tran,
Thong Minh Cao,
Phuong Nguyen Dang,
Loan Thi Hong Truong,
Nhon Van
Publication year - 2017
Publication title -
khoa học và công nghệ: tự nhiên
Language(s) - English
Resource type - Journals
ISSN - 2588-106X
DOI - 10.32508/stdjns.v1it4.476
Subject(s) - electromagnetic shielding , scanner , monte carlo method , computed tomography , nuclear medicine , normalization (sociology) , radiation protection , absorbed dose , photon , materials science , medical physics , physics , radiation , nuclear engineering , optics , medicine , engineering , radiology , mathematics , statistics , sociology , anthropology , composite material
Shielding design of Computed Tomography scanner room is a vital work to ensure radiation safety for medical physicists and public. In this paper, we measured CTDI (CT Dose Index) and determined the normalization factor between Monte Carlo simulation and experiment values of absorbed dose. Then, the absorbed dose distribution inside and outside the CT scanner room were surveyed for the variation of the room size. In addition, the influence of photons scattered from the shielding material to patients while shrinking the sizes of the room is also studied in this work.