Analisis Lactat Dehydrogenase dalam Serum Darah Menggunakan Sentrifugasi
Author(s) -
Rosnita Sebayang,
Yuana Idawati,
Hotman Sinaga
Publication year - 2020
Publication title -
jurnal keperawatan silampari
Language(s) - English
Resource type - Journals
ISSN - 2581-1975
DOI - 10.31539/jks.v4i1.1450
Subject(s) - centrifugation , lactate dehydrogenase , significant difference , chromatography , andrology , centrifuge , blood lactate , chemistry , l lactate dehydrogenase , medicine , biochemistry , enzyme , physics , blood pressure , heart rate , nuclear physics
This study shortens the laboratory service time by shortening the centrifuge time and obtaining serum as an examination material. The research method used was pre-experimental research. The results showed that the average LDH in the blood group centrifuged 3000 rpm 5 minutes was 309.9 U / L and 306.4 U / L centrifuged 4400 rpm 3 minutes in the blood group. In conclusion, there is no difference in serum LDH levels from blood centrifuged at 3000 rpm for 5 minutes and 4400 rpm for 3 minutes.
Keywords: Centrifugation, Lactate Dehydrogenase, TAT