Penentuan Nilai Faktor Mesin Pesawat Sinar-X Radiografi Digital Merek Shimadzu di RSUD Dr. Kanujoso Djatiwibowo Balikpapan
Author(s) -
Mamba'ul Fitriyana,
Suhadi Muliyono,
Kadek Subagiada
Publication year - 2020
Publication title -
progressive physics journal
Language(s) - English
Resource type - Journals
ISSN - 2722-7707
DOI - 10.30872/ppj.v1i1.563
Subject(s) - detector , digital radiography , physics , nuclear medicine , mathematics , computer science , radiography , medicine , optics , nuclear physics
The research on the determining machine factor (k) of Shimadzu digital radiograph X-ray machine have been done at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan. Machine factor should be required in the calculation of radiation exposure which produced by the X-ray machine. Measurement data that have been used as independent variables were current-time, focus distance to the detector and tube voltage variations of 80 kV to 100 kV, while the dependent variable was the S value. Determination of the k value was conducted by weighted linear regression between V2 and using the results of research by Seibert and Morin (2011) for the condition of calibrated X-ray machine. Thus, the reseach has obtained the machine factor of the Shimadzu digital radiograph X-ray machine at the RSUD Dr. Kanujoso Djatiwibowo Balikpapan.