OPTIMASI CAIRAN PEMBANGKIT MESIN PENCUCI FILM RADIOGRAF PADA LABORATORIUM FISIKA MEDIK UNNES
Author(s) -
Diah Rahayu Ningtias,
Sugeng Hadi Susilo,
S. Sutikno
Publication year - 2013
Publication title -
dinamika teknik mesin
Language(s) - English
Resource type - Journals
eISSN - 2502-1729
pISSN - 2088-088X
DOI - 10.29303/d.v3i2.78
Subject(s) - interval (graph theory) , nuclear chemistry , chemistry , materials science , physics , mathematics , combinatorics
Automatic film washing machine (TAEHN, model TM 300E in Physics Laboratory of FMIPA Unnes was successfully optimized by variating temperatures of developer and fixer solution in the range of 28˚C until 40˚C with temperature interval of 2˚C. At the first, developer solution was variated at controlled fixer temperature namely 30˚C, then it was analyzed to fixer determine density of each radiograph film. The optimum temperature of developer solution is available to varying temperature of fixer solution. The optimum temperatures of developer solution and fixer solution are 32˚C and 32˚C, respectively.