Calibration of NaI (Tl) cylindrical detector using axially shifted radioactive cylindrical sources
Author(s) -
Mona M. Gouda
Publication year - 2019
Publication title -
nuclear technology and radiation protection
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.31
H-Index - 16
eISSN - 1452-8185
pISSN - 1451-3994
DOI - 10.2298/ntrp190708035g
Subject(s) - detector , calibration , axial symmetry , radioactive source , physics , optics , photon , nuclear physics , materials science , quantum mechanics
In this article, the full energy peak efficiency of NaI detector using non-axial cylindrical sources is calculated by using a new efficient theoretical approach. This approach depends on using the efficiency transfer method and analytical calculations of the average path length of a gamma photon inside the source to the detector system. Measured efficiencies made by using 152Eu aqueous radioactive cylindrical sources with volumes 25 ml and 400 ml. Comparing calculated efficiencies to the measured one showed good agreement enabling the validation of this approach.