Simultaneous measurements of radon, thoron and equilibrium equivalent concentrations in family house - single case study
Author(s) -
Robert Lakatoš,
Sofija Forkapić,
Vanja Radolić,
Igor Čelikovíć,
Selena Samardžić,
D. Mrdja,
Kristina Bikit-Schroeder
Publication year - 2019
Publication title -
nuclear technology and radiation protection
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.31
H-Index - 16
eISSN - 1452-8185
pISSN - 1451-3994
DOI - 10.2298/ntrp190128017l
Subject(s) - radon , environmental science , quantum mechanics , physics
Simultaneous indoor measurements of radon, thoron and equilibrium equivalent con centrations, by three different continuous radon monitors in real ambient conditions, were performed and compared during two weeks period. Radon concentrations varied from 153 Bqm?3 to 870 Bqm?3 with variations of thoron concentrations in the range (2.17-219) Bqm?3. Obtained arithmetic means of equilibrium equivalent radon and thoron concentrations were (130 ? 50) Bqm?3 and (23 ? 18) Bqm?3, respectively. The inter-comparison of active radon de vices was focused on consistency of results between them selves and the influence of environmental conditions on the behavior of each detector system. The obtained statistically significant correlation between the results of used monitors, validates them for application in radon diagnostic measurements as the first step in radon remediation and dose assessment.