Extra radiation dose of lungs in male examinees during cardiac computed tomography
Author(s) -
ShinYi Tsai,
HsienChun Tseng,
Zourong Ruan,
Zumei Xu,
Yulin Liao,
Sheng-Ping ChangLai,
Chien-Yi Chen
Publication year - 2017
Publication title -
nuclear technology and radiation protection
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.31
H-Index - 16
eISSN - 1452-8185
pISSN - 1451-3994
DOI - 10.2298/ntrp1704350t
Subject(s) - computed tomography , nuclear medicine , dosimeter , medicine , tomography , thermoluminescent dosimeter , automatic exposure control , radiation dose , radiology , dosimetry
The use of multi-detector row computed tomography in cardiac examinations is increasing globally. Several hospitals are yet to establish a practical method for evaluating the extra scattering dose in the lungs (Dlung) of male examinees with different body-weights undergoing cardiac computed tomography. To measure the dose in each slice of a lung (Dlung,i), thermoluminescent dosimeters are inserted into Rando and tissue-equivalent phantoms made of polymethylmethacrylate and derived from ICRU 48 with masses from 10 to 90 kg. Dlung was evaluated by weighing the scanned volume of each slice. A practical method for determining Dlung involves 64-slice computed tomography scans using a Philips Brilliance computed tomography at 120 kV and 200 mAs, with a thickness of 1.0 mm. Dlung,i increased with distance from the scanning region. This experiment yielded Dlung values from 12.1 ? 2.1 mSv (90 kg) to 23.0 ? 3.8 mSv (10 kg). Finally, a simple equation can be used to derive the relationship between Dlung and the body-weights of a male examinee. Experimental results are compared with others in the literature.