Abstract: Every year a high number of total hip arthroplasty is reported worldwide and an increase in this number is expected. Several factors may cause hip wear, such as osteoarthritis, obesity, traffic accidents and sport practicing. Wear is a concern when considering hip prostheses, since a prosthesis presents finite life that in many cases is shorter than patient life, and leads to substitution. Also, research is constant and new developments have to be tested, which leads to the necessity of testing devices that reproduce real conditions of hip joint implant functioning. This work describes a low-cost device, according to the ISO 12242. The equipment was built, a set of three commercially available prostheses was tested and the results show wear values coherent with those found in literature. It was found a value of wear rate of (13.30 ± 3,81) mg/106 cycles; wear factor found was (0.41 ± 0,09) x 10-6 mm3 /Nm. After testing, the device was evaluated and no component presented significant wear. Keywords: Hip joint simulator; Prostheses; Wear; Arthroplasty; Test machine.