VALIDATED SPECTROPHOTOMETRIC METHOD TO DETERMINE VARDENAFIL AND SILDENAFIL IN PHARMACEUTICAL FORMS USING POTASSIUM IODIDE AND POTASSIUM IODATE
Author(s) -
Amir Alhaj Sakur,
Shaza Affas
Publication year - 2017
Publication title -
international journal of pharmacy and pharmaceutical sciences/international journal of pharmacy and pharmaceutical sciences
Language(s) - English
Resource type - Journals
eISSN - 2656-0097
pISSN - 0975-1491
DOI - 10.22159/ijpps.2017v9i11.20578
Subject(s) - molar absorptivity , potassium iodate , absorbance , vardenafil , chemistry , triiodide , potassium , iodide , potassium nitrate , iodine , dosage form , aqueous solution , nuclear chemistry , analytical chemistry (journal) , chromatography , sildenafil , inorganic chemistry , organic chemistry , medicine , physics , tadalafil , electrode , optics , electrolyte , dye sensitized solar cell
Objective: To develop and validate simple, sensitive, precise and free of organic solvents method for the determination of sildenafil (SIL) and vardenafil (VAR) in bulk and pharmaceutical formulation.Methods: The method is based on the reaction of studied drugs with a mixture of potassium iodide and potassium iodate in an aqueous medium at (25±2 °C) to form yellow coloured triiodide ions (I3-) within 45 min. The reaction is followed spectrophotometrically by measuring the absorbance at 288, 351 nm and 285, 351 nm for sildenafil and vardenafil respectively.Results: The effects of analytical parameters on the reported systems were investigated. Beer's law of SIL was obeyed in the range of (0.4-12) μg ml-1and (0.6-16) μg ml-1. Molar absorptivity was found to be (67.659 ×103) lmol/cmand (37.955×103) lmol/cm at 288 nm, 351 nm respectively. Beer's law of VAR was obeyed in the range of (0.2-13) μg/mland (0.5-40) μg/ml. Moreover, molar absorptivity’s were found to be (68.719 ×103) l mol-1 cm-1and (26.691×103) l mol-1 cm-1 at 285 nm, 351 nm respectively.Conclusion: The proposed method has been applied to determine the components in dosage forms with an average recovery of 98.15% to 103.45% and the results have been found in good agreement with those results obtained by the reference methods.