HPLC ANALYSIS OF HUMAN URINE FOR OXALATE CONTENT
Author(s) -
Mangesh V. Suryavanshi,
Swapnil Jadhav,
Rahul P. Gune,
Meenakshi Bhatia,
Yogesh S. Shouche
Publication year - 2016
Publication title -
international journal of pharmacy and pharmaceutical sciences/international journal of pharmacy and pharmaceutical sciences
Language(s) - English
Resource type - Journals
eISSN - 2656-0097
pISSN - 0975-1491
DOI - 10.22159/ijpps.2016v8i12.13168
Subject(s) - oxalate , oxalic acid , urine , chromatography , high performance liquid chromatography , chemistry , detection limit , kidney stones , kidney , acetic acid , medicine , biochemistry , organic chemistry
Objective: In the present communication, development and validation of reverse phase-high performance liquid chromatography method have been carried out for estimation of oxalate content in the urine of human volunteers with recurrent kidney stone disease and healthy status. Methods: The analysis of oxalic acid has been carried out on KYA TECH HiQ Sil C 18 HS column using a mobile phase of methanol: 0.001 N acetic acid in water (50:50, v/v) with a flow rate of 1 ml/min and detection wavelength, 237 nm. Results: Analysis of oxalate content was carried out using single point calibration method with retention at 2.705 min with good resolution parameters. Urine sample collected from kidney stone patients and healthy volunteers over the period of 24 h were analyzed and it has been found that concentration of oxalate in healthy volunteers is less than 12 µg/ml whereas that in case of kidney stone patients is in the range of 39-151 µg/ml and this data can be utilized for further interpretations about oxalate content in healthy and kidney stone diseased volunteers. This method was validated as per united states food and drug administration (USFDA) guidelines by the study of accuracy, precision, linearity, range, selectivity, the lower limit of quantitation, extraction recovery studies and stability studies for determining oxalate content in the urine of human volunteers. As relative standard deviations of oxalate content estimated are less than 5 percent, the method can be claimed accurate, precise, sensitive and selective for determining oxalate content in the urine of human volunteers. Conclusion: The results are satisfactory, proving the effectiveness of the method for analysis of oxalate content from other biological fluids with few optimizations.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here