Simultaneous LC-MS/MS Determination of Sacubitril, Valsartan and LBQ657 in Human Plasma and Its Application to Clinical Study
Author(s) -
Yufeng Ni,
Yujia Zhang,
Chong Zou,
Li Ding
Publication year - 2021
Publication title -
journal of the brazilian chemical society
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.337
H-Index - 70
eISSN - 1678-4790
pISSN - 0103-5053
DOI - 10.21577/0103-5053.20210068
Subject(s) - chromatography , valsartan , formic acid , protein precipitation , chemistry , electrospray ionization , ammonium acetate , mass spectrometry , analyte , sacubitril , selected reaction monitoring , tandem mass spectrometry , analytical chemistry (journal) , high performance liquid chromatography , medicine , blood pressure , radiology
A rapid and reproducible liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated to simultaneously determine sacubitril, valsartan and a metabolite of sacubitril (LBQ657) in human plasma using sacubitril-d4 and valsartan-d3 as the internal standards. Following protein precipitation, the analytes were operated on an Ultimate® XB-C18 column (2.1 × 50 mm, 3.5 μm, Welch) with a gradient elution with acetonitrile, and 5 mM ammonium acetate and 0.1% formic acid in water as the mobile phase. The detection was performed on a Triple Quad™ 4000 mass spectrometer coupled with an electrospray ionization source (ESI) under positive-ion multiple reaction monitoring mode. The linearities are 2.00-4000, 5.00-10000 and 5.00-10000 ng mL-1 for sacubitril, valsartan and LBQ657, respectively. The accuracy and precision of intra- and inter-day, dilution accuracy, recovery and stability of the method were all within the acceptable limits and no matrix effect or carryover was observed. The suitability of the method was successfully demonstrated in terms of the quantification of sacubitril, valsartan and LBQ657 in plasma samples collected from healthy Chinese volunteers in a clinical trial.