High performance liquid chromatographic method for the determination of guaifenesin in pharmaceutical syrups and in environmental samples
Author(s) -
Nief Rahman Ahmed,
Suhaib N. Lottfi
Publication year - 2013
Publication title -
mağallaẗ baġdād li-l-ʿulūm
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.167
H-Index - 6
eISSN - 2411-7986
pISSN - 2078-8665
DOI - 10.21123/bsj.10.3.1014-1022
Subject(s) - chromatography , calibration curve , chemistry , detection limit , acetonitrile , methanol , high performance liquid chromatography , volumetric flow rate , effluent , organic chemistry , physics , quantum mechanics , environmental engineering , engineering
A simple, precise, rapid, and accurate reversed – phase high performance liquid chromatographic method has been developed for the determination of guaifenesin in pure from pharmaceutical formulations.andindustrial effluent. Chromatography was carried out on supelco L7 reversed- phase column (25cm × 4.6mm), 5 microns, using a mixture of methanol –acetonitrile-water: (80: 10 :10 v/v/v) as a mobile phase at a flow rate of 1.0 ml.min-1. Detection was performed at 254nm at ambient temperature. The retention time for guaifenesin was found 2.4 minutes. The calibration curve was linear (r= 0.9998) over a concentration range from 0.08 to 0.8mg/ml. Limit of detection (LOD) and limit of quantification ( LOQ) were found 6µg/ml and 18µg/ml respectively. The method was validated for its linearity, precision and accuracy .The proposed method was successfully applied for the determination of guaifenesin in syrups and industrial effluent samples.