Determination of Calibration Characteristics of X-Ray Fluorescent Analysis of Rhenium
Author(s) -
Anna V. Troeglazova,
Elena V. Zlobina,
Alexander Yu. Sokolov,
Anargul K. Zhangalieva
Publication year - 2010
Publication title -
eurasian chemico-technological journal
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.227
H-Index - 9
eISSN - 2522-4867
pISSN - 1562-3920
DOI - 10.18321/ectj214
Subject(s) - rhenium , detection limit , calibration curve , calibration , fluorescence , analytical chemistry (journal) , chemistry , materials science , chromatography , optics , inorganic chemistry , mathematics , physics , statistics
Extraction concentration of rhenium by a melt of tetraoctylammonium bromide in paraffin was used with the aim of preparation of calibration samples for roentgen fluorescent analysis. The calibration characteristics of the roentgen fluorescent analysis of rhenium is established, its linearity is evaluated, a limit of detection and a limit of quantification (determination) of rhenium are calculated. The maximal value of the relative total standard uncertainty of the calibration characteristic does not exceed 1.45 %.