Areas of effectiveness of half-sine monophasic and biphasic depolarizing defibrillation pulses on the diagram of energy / phase of fibrillation cycle
Author(s) -
B. B. Gorbunov,
Iryesterenko,
D. V. Telyshev,
С. В. Селіщев
Publication year - 2021
Publication title -
ural radio engineering journal
Language(s) - English
Resource type - Journals
eISSN - 2588-0462
pISSN - 2588-0454
DOI - 10.15826/urej.2021.5.4.003
Subject(s) - defibrillation , ventricular fibrillation , pulse (music) , fibrillation , energy (signal processing) , cardiology , medicine , materials science , computer science , physics , atrial fibrillation , optics , quantum mechanics , detector
The aim of this study is to compare the areas of effectiveness of half-sine monophasic and biphasic depolarizing defibrillation pulses in the diagram of energy / phase of fibrillation cycle. The study was carried out on the ten Tusscher-Panfilov 2006 model of the human ventricular myocyte under the influence of simulated fibrillation in the BeatBox simulation environment under the Fedora operating system. The simulation was carried out on a computer under the Windows 10 operating system, the Fedora operating system was implemented in the Oracle VM VirtualBox virtualization environment. The results of computer simulations have shown that the areas of effectiveness for monophasic and biphasic defibrillation pulses are significantly different. In a biphasic pulse, the fraction of a fibrillation cycle at which refractoriness is extended is significantly higher than that of a monophasic pulse at low defibrillation pulse energies. It can be assumed that this provides the energy advantage of a biphasic defibrillation pulse over a monophasic one.