z-logo
open-access-imgOpen Access
Modular Necks and Corrosion - Review of Five Cases
Author(s) -
Timothy McTighe,
Declan Brazil
Publication year - 2016
Publication title -
reconstructive review
Language(s) - English
Resource type - Journals
eISSN - 2331-2270
pISSN - 2331-2262
DOI - 10.15438/rr.6.3.163
Subject(s) - fretting , implant , medicine , surgery , modular design , total hip arthroplasty , soft tissue , dentistry , materials science , metallurgy , computer science , operating system
This report updates previous articles and commentary presented on Modular Necks from our Tissue Sparing Implant Study Group. In July 2012 we reported that at two years post-op we had encountered no modular neck taper failures or any signs of fretting corrosion, or pseudo tumors associated with the ARC™ Stem.Today we describe five patients out of five hundred and forty-two who had total hip arthroplasty revision [titanium alloy stem, cobalt-chromium modular neck (c.c.) and c.c. modular head (32 mm or 36 mm), highly-cross-linked polyethylene liner, metal titanium plasma sprayed cementless metal cup]. All patients’ were female and all demonstrated progressive hip pain or late instability. All  had debridement of the periarticular soft tissue, stem extraction with new primary length cementless stem replacement. At revision and early follow up all patients are doing well, however, we recommend heighten awareness in all active female patients with modular neck stem junctions.