Implantation of an S-ICD in a Patient with a DDD Pacemaker and Congenitally Corrected Transposition of the Great Arteries
Author(s) -
Yu Zhang,
Wenlong Dai,
ChungYing Lin,
Qiaoyuan Li,
Chengjun Guo
Publication year - 2021
Publication title -
cardiovascular innovations and applications
Language(s) - English
Resource type - Journals
eISSN - 2009-8782
pISSN - 2009-8618
DOI - 10.15212/cvia.2019.0597
Subject(s) - medicine , great arteries , cardiology , atrioventricular block , sudden cardiac death , defibrillation , heart disease , implantable cardioverter defibrillator
Subcutaneous implantable cardioverter-defibrillator (S-ICD) therapy has become a viable alternative to conventional transvenous ICD implantation. Patients with congenitally corrected transposition of the great arteries (ccTGA) have a high risk of sudden cardiac death due to malignant arrhythmia. The interaction between the S-ICD system and the transvenous pacemaker system is not fully understood. We report a case of S-ICD implantation in a patient with ccTGA and a DDD pacemaker. The patient was a 30-year-old man with a previously placed pacemaker with diagnoses of congenital heart disease, ccTGA (SLL), left atrioventricular valve insufficiency, and third-degree atrioventricular block. He presented with an out-of-hospital cardiac arrest, and an S-ICD was implanted to prevent sudden cardiac death. Defibrillation checks were performed successfully. We tested the compatibility of the DDD pacemaker with the S-ICD and found that there was no interference between them. In conclusion, an S-ICD system is a reasonable and safe option in patients with ccTGA.