Low cost multifunctional 3D printed image quality and dose verification phantom for an image-guided radiotherapy system
Author(s) -
Jia-Rong Wu,
MinTeh Yu,
Shih-Han Chen,
Shu-Hsien Liao,
Yu Jen Wang
Publication year - 2022
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0266604
Subject(s) - imaging phantom , quality assurance , linear particle accelerator , image guided radiation therapy , calibration , image quality , medical physics , image resolution , computer science , hounsfield scale , biomedical engineering , medical imaging , consistency (knowledge bases) , dosimetry , materials science , nuclear medicine , optics , beam (structure) , artificial intelligence , medicine , physics , computed tomography , radiology , image (mathematics) , external quality assessment , pathology , quantum mechanics
Purpose Image-guided radiation therapy (IGRT) is used to precisely deliver radiation to a tumour to reduce the possible damage to the surrounding normal tissues. Clinics use various quality assurance (QA) equipment to ensure that the performance of the IGRT system meets the international standards set for the system. The objective of this study was to develop a low-cost and multipurpose module for evaluating image quality and dose. Methods A multipurpose phantom was designed to meet the clinical requirements of high accuracy, easy setup, and calibration. The outer shell of the phantom was fabricated using acrylic. Three dimensional (3D) printing technology was used to fabricate inner slabs with the characteristics of high spatial resolution, low-contrast detectability, a 3D grid, and liquid-filled uniformity. All materials were compatible with magnetic resonance (MR). Computed tomography (CT) simulator and linear accelerator (LINAC) modules were developed and validated. Results The uniformity slab filled with water is ideal for the assessment of Hounsfield units, whereas that filled with wax is suitable for consistency checks. The high-spatial-resolution slab enables measurements with a resolution up to 5 lp/cm. The low-contrast detectability slab contains rods of 5 different sizes that can be clearly visualised. These components meet the American College of Radiology (ACR) standards for QA of CT simulators and LINACs. Conclusions The multifunctional phantom module meets the ACR recommended QA guidelines and is suitable for both LINACs and CT-sim. Further measurements in an MR simulator and an MR linear accelerator (MR-LINAC) will be arranged in the future.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here