z-logo
open-access-imgOpen Access
Source apportionment and quantification of liquid and headspace leaks from closed system drug-transfer devices via Selected Ion Flow Tube Mass Spectrometry (SIFT-MS)
Author(s) -
Amos Doepke,
Robert Streicher
Publication year - 2021
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0258425
Subject(s) - chromatography , mass spectrometry , detection limit , volatile organic compound , chemistry , analytical chemistry (journal) , leak , environmental science , organic chemistry , environmental engineering
A system to differentiate and quantify liquid and headspace vapor leaks from closed system drug-transfer devices (CSTDs) is presented. CSTDs are designed to reduce or eliminate hazardous drug (HD) exposure risk when compounding and administering HDs. CSTDs may leak liquid, headspace, or a mixture of the two. The amount of HD contained in liquid and headspace leaks may be substantially different. Use of a test solution containing two VOCs with differences in ratios of VOC concentrations in the headspace and liquid enables source apportionment of leaked material. SIFT-MS was used to detect VOCs from liquid and headspace leaks in the vapor phase. Included in this report is a novel method to determine the origin and magnitude of leaks from CSTDs. A limit of leak detection of 24 μL of headspace vapor and 0.14 μL of test liquid were found using Selected Ion Flow Tube Mass Spectrometry (SIFT-MS).