Longitudinal stability of a multimodal visco-elastic polyacrylamide gel phantom for magnetic resonance and ultrasound shear-wave elastography
Author(s) -
Masashi Usumura,
Riwa Kishimoto,
Koki Ishii,
Eika Hotta,
Jeff Kershaw,
Tatsuya Higashi,
Takayuki Obata,
Mikio Suga
Publication year - 2021
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0250667
Subject(s) - imaging phantom , magnetic resonance elastography , elastography , viscoelasticity , materials science , biomedical engineering , magnetic resonance imaging , repeatability , ultrasound , stiffness , shear modulus , polyacrylamide , nuclear magnetic resonance , nuclear medicine , composite material , physics , acoustics , radiology , medicine , chemistry , chromatography , polymer chemistry
We evaluated the long-term stability of a newly developed viscoelastic phantom made of polyacrylamide (PAAm) gel for magnetic resonance elastography (MRE) and ultrasound-based shear-wave elastography (US SWE). The stiffness of the cylindrical phantom was measured at 0, 13 and 18 months. Storage and loss moduli were measured with MRE, and shear-wave speed (SWS) was measured with US SWE. Long-term stability was evaluated in accordance with the Quantitative Imaging Biomarker Alliance (QIBA) profiles for each modality. The initial storage and loss moduli of the phantom were 5.01±0.22 and 1.11±0.15 respectively, and SWS was 2.57±0.04 m/s. The weight of the phantom decreased by 0.6% over the 18 months. When measured with MRE, the stiffness of the phantom decreased and changes to the storage and loss moduli were -3.0% and -4.6% between 0 and 13 months, and -4.3% and 0.0% between 0 and 18 months. The US measurements found that SWS decreased by 2.4% over the first 13 months and 3.6% at 18 months. These changes were smaller than the tolerances specified in the QIBA profiles, so the viscoelastic PAAm gel phantom fulfilled the condition for long-term stability. This new phantom has the potential to be used as a quality assurance and quality control phantom for MRE and US SWE.