Development and validation of a stability indicating RP-HPLC-DAD method for the determination of bromazepam
Author(s) -
Hany W. Darwish,
Nesma A. Ali,
Ibrahim A. Naguib,
Mohamed Mohamady Ghobashy,
Abdullah M. Al-Hossaini,
Maha M. Abdelrahman
Publication year - 2021
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0244951
Subject(s) - bromazepam , chromatography , high performance liquid chromatography , forced degradation , detection limit , chemistry , calibration curve , elution , methanol , dosage form , analytical chemistry (journal) , biochemistry , receptor , organic chemistry , benzodiazepine
A reliable, selective and sensitive stability-indicating RP-HPLC assay was established for the quantitation of bromazepam (BMZ) and one of the degradant and stated potential impurities; 2-(2-amino-5-bromobenzoyl) pyridine (ABP). The assay was accomplished on a C 18 column (250 mm × 4.6 mm i.d., 5 μm particle size), and utilizing methanol-water (70: 30, v/v) as the mobile phase, at a flow rate of 1.0 ml min -1 . HPLC detection of elute was obtained by a photodiode array detector (DAD) which was set at 230 nm. ICH guidelines were adhered for validation of proposed method regarding specificity, sensitivity, precision, linearity, accuracy, system suitability and robustness. Calibration curves of BMZ and ABP were created in the range of 1–16 μg mL -1 with mean recovery percentage of 100.02 ± 1.245 and 99.74 ± 1.124, and detection limit of 0.20 μg mL -1 and 0.24 μg mL -1 respectively. BMZ stability was inspected under various ICH forced degradation conditions and it was found to be easily degraded in acidic and alkaline conditions. The results revealed the suitability of the described methodology for the quantitation of the impurity (ABP) in a BMZ pure sample. The determination of BMZ in pharmaceutical dosage forms was conducted with the described method and showed mean percentage recovery of 99.39 ± 1.401 and 98.72 ± 1.795 (n = 6), respectively. When comparing the described procedure to a reference HPLC method statistically, no significant differences between the two methods in regard to both accuracy and precision were found.