Evaluation of biomechanically corrected intraocular pressure using Corvis ST and comparison of the Corvis ST, noncontact tonometer, and Goldmann applanation tonometer in patients with glaucoma
Author(s) -
Yoshitaka Nakao,
Yoshiaki Kiuchi,
Hideaki Okumichi
Publication year - 2020
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0238395
Subject(s) - intraocular pressure , scheimpflug principle , goldmann applanation tonometer , medicine , ophthalmology , glaucoma , cornea
Purpose The aim of the study was to investigate the effects of various anatomical structures on intraocular pressure (IOP) measurements obtained by the Corneal Visualization Scheimpflug Technology (Corvis ST), Goldmann applanation tonometer (GAT), and noncontact tonometer (NCT), as well as to assess the interchangeability among the four types of IOP measurement: IOP-GAT, IOP-NCT, IOP-Corvis, and biomechanically corrected IOP (bIOP-Corvis), with a particular focus on bIOP-Corvis. Materials and methods We included 71 patients with primary open-angle glaucoma and assessed their IOP measurements obtained with the GAT, NCT, and Corvis ST using a repeated measures ANOVA, a paired t-test with Bonferroni correction, stepwise multiple regression analyses and Bland–Altman plots. Results IOP-GAT showed the highest values (13.5 ± 2.1 mmHg [mean ± standard deviation]), followed by IOP-NCT (13.2 ± 2.7 mmHg), IOP-Corvis (10.6 ± 2.8 mmHg), and bIOP-Corvis (10.0 ± 2.3 mmHg). With exceptions of bIOP-Corvis and IOP-GAT, all IOP variations were explained by regression coefficients involving the central corneal thickness. Bland–Altman plots showed a mean difference between IOP-GAT and the other IOP measurements (IOP-Corvis, bIOP-Corvis, and IOP-NCT), which were -2.90, -3.48, and -0.29 mmHg, respectively. The widths of the 95% limits of agreement between all pairs of IOP measurements were greater than 3 mmHg. Conclusion IOP values obtained with the Corvis ST, NCT, and GAT were not interchangeable. The bIOP-Corvis measurement corrected for the ocular structure.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here