Incubator-based Sound Attenuation: Active Noise Control In A Simulated Clinical Environment
Author(s) -
George M. Hutchinson,
Lilin Du,
Kaashif A. Ahmad
Publication year - 2020
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0235287
Subject(s) - incubator , noise (video) , microphone , octave band , acoustics , sound pressure , neonatal intensive care unit , active noise control , alarm , background noise , noise reduction , sound level meter , sound (geography) , medicine , octave (electronics) , computer science , engineering , noise level , electrical engineering , physics , artificial intelligence , pediatrics , microbiology and biotechnology , image (mathematics) , biology
Objective Noise in the neonatal intensive care unit can be detrimental to the health of the hospitalized infant. Means of reducing that noise include staff training, warning lights, and ear coverings, all of which have had limited success. Single family rooms, while an improvement, also expose the hospitalized infant to the same device alarms and mechanical noises found in open bay units. Methods We evaluated a non-contact incubator-based active noise control device (Neoasis™, Invictus Medical, San Antonio, Texas) in a simulated neonatal intensive care unit (NICU) setting to determine whether it could effectively reduce the noise exposure of infants within an incubator. In the NICU simulation center, we generated a series of clinically appropriate sound sequences with bedside medical devices such as a patient monitor and fluid infusion devices, hospital air handling systems, and device mechanical sounds. A microphone-equipped infant mannequin was oriented within an incubator. Measurements were made with the microphones with the Neoasis™ deactivated and activated. Results The active noise control device decreased sound pressure levels for certain alarm sounds by as much as 14.4 dB (a 5.2-fold reduction in sound pressure) at the alarm tone’s primary frequency. Frequencies below the 2 kHz octave band were more effectively attenuated than frequencies at or above the 2 kHz octave band. Background noise levels below 40 dBA were essentially not impacted by the active noise control device. Conclusions The active noise control device further reduces noise inside infant incubators. Device safety and potential health benefits of the quieter environment should be verified in a clinical setting.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here