The Quansys multiplex immunoassay for serum ferritin, C-reactive protein, and α-1-acid glycoprotein showed good comparability with reference-type assays but not for soluble transferrin receptor and retinol-binding protein
Author(s) -
Razieh Esmaeili,
Ming Zhang,
Maya Sternberg,
Carine Mapango,
Christine M. Pfeiffer
Publication year - 2019
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0215782
Subject(s) - soluble transferrin receptor , retinol binding protein 4 , transferrin , ferritin , orosomucoid , nephelometry , retinol binding protein , immunoassay , multiplex , chemistry , microbiology and biotechnology , biology , glycoprotein , biochemistry , medicine , immunology , endocrinology , antibody , iron deficiency , retinol , vitamin , anemia , iron status , adipokine , insulin resistance , bioinformatics , insulin
The Quansys multiplex (Q-Plex) measures ferritin (Fer), soluble transferrin receptor (sTfR), C-reactive protein (CRP), α-1-acid glycoprotein (AGP), and retinol-binding protein (RBP). We compared Q-Plex results with reference-type assays and evaluated Q-Plex performance. Pearson correlation and Lin’s concordance coefficients between the Q-Plex and reference assay were: Fer 0.98 and 0.91, sTfR 0.88 and 0.35, CRP 0.98 and 0.98, AGP 0.82 and 0.81, and RBP 0.68 and 0.31, respectively. The median relative difference between the Q-Plex and reference assay were: Fer -2.4%, sTfR 107%, CRP 0.03%, AGP -1.3%, and RBP 51%. The Q-Plex intra-assay CVs were <5%; the inter-assay CVs were higher: Fer 11%, sTfR 14%, CRP 9.3%, AGP 7.5%, and RBP 19%. EDTA plasma produced 74% higher Q-Plex sTfR concentrations compared to serum. Analyte stability was good for ≤5 freeze-thaw cycles. After adjusting Q-Plex data to the reference assays, sensitivity and specificity were >85% for Fer and CRP; specificity was >85% for sTfR, AGP, and RBP. Using performance criteria derived from biologic variation, Fer, CRP, and AGP met the minimum allowable imprecision (<10.7%, <31.7%, and <8.5%, respectively) and difference from the reference assay (<±7.7%, <±32.7%, and <±10.3%, respectively), while sTfR and RBP exceeded these thresholds (<8.5% and <7.8% for imprecision and <±7.7% and <±12% for difference, respectively). The Q-Plex measures multiple biomarkers simultaneously, is easy to perform, and uses small sample volumes. With some improvements in accuracy and precision (i.e., sTfR and RBP), this assay could be a useful tool for low-resource laboratories conducting micronutrient surveys for epidemiologic screening applications. These findings need to be verified using other populations, particularly those with inadequate micronutrient status.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here