Dosimetric characteristics of a reusable 3D radiochromic dosimetry material
Author(s) -
Jong Min Park,
SoYeon Park,
Cheol Young Choi,
Minsoo Chun,
Ji Hye Han,
Jin Dong Cho,
Jung-In Kim
Publication year - 2017
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0180970
Subject(s) - dosimetry , medical physics , nuclear medicine , medicine
Purpose To investigate the dosimetric characteristics of PRESAGE REU dosimeters. Methods Commercially available PRESAGE REU dosimeters (size of 10 mm × 10 mm × 45 mm) were divided into two groups, with one of the groups placed at room temperature of 22°C (RT group) and another group placed at low temperature of 10°C (LT group). A total of 3 dosimeters (set of dosimeters) were irradiated at a time, with doses of 1 Gy, 2 Gy, 4 Gy, 8 Gy, 12 Gy, 16 Gy, and 20 Gy, at a nominal dose rate of 400 MU/min at temperature of 22°C. The dosimeters were irradiated three additional times by delivering the same doses as those during the initial irradiations (4 irradiation cycles). Optical density (OD) was assessed using optical CT scanning. Results Considering both linearity and sensitivity of the OD curves, R 2 above 0.95 and sensitivity above 0.04 ΔOD/Gy were observed at the 1 st irradiation (reading time ≤ 6 h) and 2 nd irradiation (reading time = 0.5 h) for the RT group. For the LT group, those values were observed at the 1 st irradiation (reading time ≤ 2 h), and the 3 rd and 4 th irradiations (both reading times = 0.5 h). Considering the reproducibility of signals in response to the same dose, dosimeters in the RT group showed average deviations among dosimeters less than 5% (the 1 st and 2 nd irradiations at the reading time of 0.5 h), while for dosimeters in the LT group showed average deviations among dosimeters less than 6% (the 3 rd and 4 th irradiations at the reading time of 0.5 h). For the rest, the OD curves were not linear, sensitivities of the dosimeters were lower than 0.04 ΔOD/Gy, and OD deviations at the same dose were larger than 6%. Conclusions At room temperature, PRESAGE REU dosimeters could be used for dose measurement only for up to two dose measurement sessions. At low temperatures, usage of PRESAGE REU dosimeters for dose measurement seems to be possible from the 3 rd irradiation. When reusing PRESAGE REU dosimeters, the OD curve should be re-defined for every measurement session because the shape of this curve depends on the irradiation history.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here