Do We Deliver the Pressures We Intend to When Using a T-Piece Resuscitator?
Author(s) -
Evelien Roegholt,
Jeroen J van Vonderen,
Frans J. Walther,
Charles Christoph Roehr,
Arjan B. te Pas
Publication year - 2013
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0064706
Subject(s) - medicine , peak inspiratory pressure , anesthesia , respiratory system , tidal volume , anatomy
Background A T-piece resuscitator (TPR) uses a built-in manometer to set the inflation pressures, but we are not informed what pressures are actually delivered distally. Aim of this study was to measure the proximal and distal pressures under different gas conditions when using a TPR. Methodology/Findings A test lung was ventilated using a TPR (PIP 25 cmH 2 O, PEEP 5 cmH 2 O) with a gas flow rate of 8 L/min. A) Pressure delivered by six different TPRs was tested. To test variability 20 participants were asked to set PEEP and PIP pressures to 25/5 cmH 2 O. B) PIP and PEEP were measured proximal and distal of the TPR when using standard tubing or heated tubing with or without a humidifier. In experiment A mean (SD) proximal PIP and PEEP of the TPRs were respectively 20.3 (0.3) cmH 2 O (19.9–20.6 cmH 2 O) and 4.9 (0.1) cmH 2 O. When 20 participants set pressures; PIP 26.7 (0.5) cm H 2 O and PEEP 5.9 (0.44) cmH 2 O were measured. Experiment B showed that the decrease of PIP between proximal and distal pressures was not clinically significant. However there was a significant decrease of PEEP using the standard tubing (5.1 (0.1) cmH 2 O proximally versus 4.8 (0.2) cmH 2 O distally; p<0.001) compared to, when using a humidifier with associated tubing and the humidifier turned on, 5.1 (0.1) proximally versus 3.9 (0.2) cmH 2 O distally; (p<0.001). Conclusion/Significance The accuracy of the built-in manometer of a TPR is acceptable. Most pressures set proximally are comparable to the actual pressures delivered distally. However, when using tubing associated with the humidifier PEEP decreases distally by 1.1–1.2 cmH 2 O and users should anticipate on this.