Performance of Three LED-Based Fluorescence Microscopy Systems for Detection of Tuberculosis in Uganda
Author(s) -
H Albert,
Yukari C. Manabe,
George Lukyamuzi,
Patrick Ademun,
Sheena Mukkada,
Barnabas Nyesiga,
Moses Joloba,
C. N. Paramasivan,
Mark D. Perkins
Publication year - 2010
Publication title -
plos one
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.99
H-Index - 332
ISSN - 1932-6203
DOI - 10.1371/journal.pone.0015206
Subject(s) - microscopy , tuberculosis , mycobacterium tuberculosis , medicine , fluorescence microscope , sputum , human immunodeficiency virus (hiv) , staining , pathology , fluorescence , virology , optics , physics
Background Direct smear microscopy using Ziehl-Neelsen (ZN) staining is the mainstay of tuberculosis (TB) diagnosis in most high burden countries, but is limited by low sensitivity in routine practice, particularly in high human immunodeficiency virus (HIV) prevalence settings. Methods We compared the performance of three commercial light emitting diode (LED)-based microscopy systems (Primostar™ iLED, Lumin™ and AFTER®) for fluorescent detection of Mycobacterium tuberculosis with ZN microscopy on slides prepared from sputum of TB suspects. Examination time for LED-based fluorescent microscopy (LED FM) and ZN slides was also compared, and a qualitative user appraisal of the LED FM systems was carried out. Results LED FM was between 5.6 and 9.4% more sensitive than ZN microscopy, although the difference was not statistically significant. There was no significant difference in the sensitivity or specificity of the three LED FM systems, although the specificity of Fraen AFTER was somewhat lower than the other LED FM methods. Examination time for LED FM was 2 and 4 times less than for ZN microscopy. LED FM was highly acceptable to Ugandan technologists, although differences in operational performance of the three systems were reported. Conclusions LED FM compares favourably with ZN microscopy, with equivalent specificity and a modest increase in sensitivity. Screening of slides was substantially quicker using LED FM than ZN, and LED FM was rated highly by laboratory technologists. Available commercial systems have different operational characteristics which should be considered prior to programmatic implementation.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here