z-logo
open-access-imgOpen Access
Onboard absolute radiometric calibration and validation of the satellite calibration spectrometer on HY-1C
Author(s) -
Heyu Xu,
Liming Zhang,
Wenjiang Huang,
W. Xu,
Xiaolong Si,
Hongyao Chen,
Xin Li,
Qingjun Song
Publication year - 2020
Publication title -
optics express
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.394
H-Index - 271
ISSN - 1094-4087
DOI - 10.1364/oe.402616
Subject(s) - remote sensing , radiometric calibration , calibration , spectroradiometer , bidirectional reflectance distribution function , radiometry , satellite , imaging spectrometer , environmental science , moderate resolution imaging spectroradiometer , optics , spectrometer , spectral bands , irradiance , reflectivity , geology , physics , quantum mechanics , astronomy
As the reference radiometric calibration standard of sensors on the Haiyang-1C (HY-1C) satellite platform, the satellite calibration spectrometer (SCS) is equipped with an onboard calibration system composed of double solar diffusers and an erbium-doped diffuser to monitor the postlaunch radiometric response change. Herein, through onboard calibration data analysis, the calibration diffuser performance remains stable without degradation, and the Moderate Resolution Imaging Spectroradiometer (MODIS) on Terra is adopted as a reference to repeatedly verify onboard radiometric calibration results by selecting different dates and reflectance scenes. The SCS equivalent reflectance is obtained by combining the mean digital number (DN) of the SCS crossing area image with the radiometric calibration coefficient. The spectral reflectance is obtained via interpolation and iteration, which is adopted as the actual MODIS incident pupil spectral reflectance because the small imaging time interval can be ignored and almost vertically observed, and it is convoluted with the MODIS spectral response function to obtain the predicted equivalent reflectance. Validation is completed by comparing the predicted MODIS equivalent reflectance to the measured value based on the onboard calibration coefficient. The results show that (1) the difference between the measured and predicted MODIS band equivalent reflectance is between -0.00466 and 0.0039, and (2) the percentage difference between the measured and predicted MODIS band equivalent reflectance ranges from 4.17% and 1.24%, indicating that the calibration system carried on HY-1C can perform high-precision SCS radiometric calibration, meeting the cross-calibration accuracy requirements of other loads on the same platform.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here