Eye motion correction algorithm for OCT-based corneal topography
Author(s) -
Elias Pavlatos,
David Huang,
Yan Li
Publication year - 2020
Publication title -
biomedical optics express
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.362
H-Index - 86
ISSN - 2156-7085
DOI - 10.1364/boe.412209
Subject(s) - scheimpflug principle , corneal topography , keratometer , optics , astigmatism , repeatability , zernike polynomials , computer science , cornea , computer vision , artificial intelligence , physics , mathematics , wavefront , statistics
With its sequential image acquisition, OCT-based corneal topography is often susceptible to measurement errors due to eye motion. We have developed a novel algorithm to detect eye motion and minimize its impact on OCT topography maps. We applied the eye motion correction algorithm to corneal topographic scans acquired using a 70 kHz spectral-domain OCT device. OCT corneal topographic measurements were compared to those from a rotating Scheimpflug camera topographer. The motion correction algorithm provided a 2-4 fold improvement in the repeatability of OCT topography and its agreement with the standard Scheimpflug topographer. The repeatability of OCT Zernike-based corneal mean power, cardinal astigmatism, and oblique astigmatism after motion detection was 0.14 D, 0.28 D, and 0.24 D, respectively. The average differences between the two devices were 0.19 D for simulated keratometry-based corneal mean power, 0.23 D for cardinal astigmatism, and 0.25 D for oblique astigmatism. Our eye motion detection method can be applied to any OCT device, and it therefore represents a powerful tool for improving OCT topography.