Point-of-Care Fibrinogen Testing in Pregnancy
Author(s) -
Daniel Katz,
Joshua Hamburger,
Dorian Batt,
Jeffrey D. Zahn,
Yaakov Beilin
Publication year - 2019
Publication title -
anesthesia and analgesia/anesthesia and analgesia
Language(s) - Uncategorized
Resource type - Journals
SCImago Journal Rank - 1.404
H-Index - 201
eISSN - 1526-7598
pISSN - 0003-2999
DOI - 10.1213/ane.0000000000003301
Subject(s) - thromboelastography , fibrinogen , medicine , point of care testing , mean difference , obstetrics , population , significant difference , pregnancy , confidence interval , platelet , biology , immunology , environmental health , genetics
Agreement between estimated fibrinogen concentration via thromboelastography and traditional assays is not established in the parturient. We therefore recruited 56 parturients and performed Clauss and functional fibrinogen level (FLEV) tests. Mean difference of measurements was 36.8 mg/dL (95% CI, 21.8-51.9) with a standard deviation of 52.8 mg/dL. Calculated limits of agreement were 140.2 mg/dL (95% CI, 166.3-114.6) and -66.6 mg/dL (95% CI, -40.8 to -92.5), within the maximum allowable difference of 165 mg/dL. We therefore conclude that while most measurements fell within the limits of agreement, more work is needed to clearly define the role of this test in the obstetric population.