In vitro assessment of inter and intra batch variability of a breath‐enhanced jet nebulizer
Author(s) -
Rehman Mariam,
Hurren Antony,
Metcalf Adam,
Hatley Ross
Publication year - 2013
Publication title -
clinical and translational allergy
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.979
H-Index - 37
ISSN - 2045-7022
DOI - 10.1186/2045-7022-3-s1-p11
Subject(s) - nebulizer , medicine , chromatography , particle size , salbutamol , reproducibility , aerosolization , volumetric flow rate , coefficient of variation , biomedical engineering , particle size distribution , aerosol , washout , inhalation , anesthesia , chemistry , asthma , organic chemistry , physics , quantum mechanics
Background Guidelines recommend that patients presenting with acute severe asthma with life-threatening features receive the necessary high doses of b2-agonists via the nebulized route.The SideStream Plus (SS+; Philips Respironics) is a breath-enhanced nebulizer designed to maximize respirable output, while minimizing treatment time [1]. Prior to the introduction of a new version of nebulizer, various in vitro assessments, including inter and intra batch variability, are performed using a limited number of devices. We present results of an analysis of inter and intra batch variability of SS+ nebulizers. Method SS+ nebulizers (3 batches; n=30 per batch) were assessed in terms of 5 experimental parameters; nebulization time, residual volume, particle size distribution (mass median diameter (MMD), fine particle fraction (FPF)), and emitted dose. Each batch was provided with a driving flow from a different Porta-neb compressor (Philips Respironics). Measurement of particle size distribution was achieved using a laser diffractor (Malvern Spraytec) with an extraction flow of 15 l/min. Each nebulizer was weighed and filled with 2.5 ml salbutamol sulphate (albuterol sulfate) solution (1 mg/ml). To assess nebulization duration the nebulizers were run continuously until ‘end of treatment’, defined as when the obscuration level of the sample in the laser diffractor fell below 5% for 5 s. At this point the SS+ were re-weighed to calculate residual volume. The dose delivered to a filter placed between the laser diffractor and extraction air flow was quantified using high performance liquid chromatography. The coefficient of variation (CV) for each experimental parameter was calculated for each batch.