Comparison of Biomechanical Failure Loads Between Tape-Type and Conventional Sutures in Internal Knotless Anchor–Based Constructs
Author(s) -
HaoChun Chuang,
Joe-Zhi Yen,
Chih-Kai Hong,
Kai-Lan Hsu,
Fa-Chuan Kuan,
Yueh Chen,
HaoMing Chang,
Wei-Ren Su
Publication year - 2022
Publication title -
orthopaedic journal of sports medicine
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 1.329
H-Index - 35
ISSN - 2325-9671
DOI - 10.1177/23259671211072523
Subject(s) - clamping , fibrous joint , slippage , medicine , tension (geology) , spiral (railway) , interlocking , washer , biomedical engineering , surgery , orthodontics , structural engineering , composite material , ultimate tensile strength , materials science , engineering , mechanical engineering
Background: Despite the increasing prevalence of tape-type sutures, whether internal knotless anchors can consistently affix tape-type sutures has not been thoroughly investigated.Purpose: To evaluate whether substituting tape-type sutures for conventional sutures influences the suture-holding strength of internal knotless anchors.Study Design: Controlled laboratory study. Level of evidence, 5.Methods: A total of 3 internal knotless anchors were tested: a spiral core clamping anchor (Footprint Ultra PK), a winged clamping anchor (PopLok), and a spooling anchor (ReelX STT). Four constructs were compared for each type of anchor, with the anchor double or quadruple loaded with tape-type sutures or conventional sutures. The testing protocol comprised preloading suture tension to 10 N; cyclic loading, in which tension increased in increments of 10 N from 10 to 90 N; and a load-to-failure stage set at a speed of 0.5 mm/s. The clinical failure load (CFL) was defined as suture slippage of ≥3 mm. Also, 1-way analysis of variance and power analysis were used to compare the CFLs of the constructs.Results: For the quadruple-loaded spiral core clamping anchors, a significant reduction in CFLs was seen with conventional sutures over tape-type sutures (138.10 ± 4.73 vs 80.00 ± 12.25 N, respectively; P < .001). This reduction was not observed under the double-loaded condition (conventional vs tape type: 76.00 ± 5.48 vs 80.00 ± 10.00 N, respectively). Substitution of the suture materials did not significantly reduce the CFLs for the winged clamping anchors (conventional vs tape type: 40.00 ± 10.00 vs 30.00 ± 7.07 N for double loaded, respectively, and 64.00 ± 13.41 vs 50.00 ± 10.00 N for quadruple loaded, respectively) or the spooling anchors (conventional vs tape type: 62.00 ± 19.23 vs 56.32 ± 20.20N for double loaded, respectively, and 72.00 ± 21.68 vs 84.00 ± 13.42 N for quadruple loaded, respectively).Conclusion: Substituting tape-type sutures for conventional sutures increased the CFLs of some internal knotless anchors. With specific suture-anchor combinations, quadruple-loaded conventional suture anchors had CFLs higher than those of double-loaded conventional suture anchors.Clinical Relevance: When multiple tape-type sutures are used in conjunction with a clamping anchor, clinicians should note a possible reduction in CFLs and resultant early suture slippage.

The content you want is available to Zendy users.

Already have an account? Click here to sign in.
Having issues? You can contact us here