z-logo
Premium
A prospective study of complications in Hickman right‐atrial catheters in marrow transplant patients
Author(s) -
Ulz L,
Petersen FB,
Ford R,
Blakely W,
Bennett C,
Grimm M,
Hickman RO
Publication year - 1990
Publication title -
journal of parenteral and enteral nutrition
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.935
H-Index - 98
eISSN - 1941-2444
pISSN - 0148-6071
DOI - 10.1177/014860719001400127
Subject(s) - medicine , catheter , surgery , bone marrow transplantation , bone marrow transplant , bone marrow , urokinase , transplantation
One hundred forty‐three large bore right‐atrial catheters (RACs) in 111 bone marrow transplant recipients were prospectively analyzed for the incidence of complications interfering with catheter function. Of the 143 RACs, 108 (76%) were patent on departure from Seattle or death. Infectious complications occurred in 63 (44%) of the catheters, resulting in 18 (13%) removals. Noninfectious complications occurred with 54 (38%) of the catheters, resulting in 17 (12%) removals. Successful interventions included catheter repair, dissolving precipitates with ammonium chloride solution, and declotting catheter lumens with heparin and/or urokinase. We conclude that RACs continue to be reliable and safe devices to use in marrow transplant patients. Complications interfering with catheter function are frequent; however, most can be resolved without the need to remove the catheter.