Clinical evaluation of the Lumac bioluminescence method for screening urine specimens
Author(s) -
Elaine Martin,
Jonathan Côté,
Libby Perry,
William J. Martin
Publication year - 1985
Publication title -
journal of clinical microbiology
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 2.349
H-Index - 255
eISSN - 1070-633X
pISSN - 0095-1137
DOI - 10.1128/jcm.22.1.19-22.1985
Subject(s) - urine , bioluminescence , predictive value , biology , microbiology and biotechnology , medicine , chromatography , chemistry , ecology
A bioluminescence method for screening urine cultures to provide rapid reporting of negative specimens and to select appropriate urine cultures for direct application of automated identification methods was evaluated. A total of 2,000 specimens were processed in the Lumac Biocounter (3M, St. Paul, Minn.), and the results were compared with quantitative culture techniques by using a 0.001-ml inoculating loop. A total of 841 specimens were positive by the bioluminescence method; 291 specimens were culture positive (greater than or equal to 50,000 CFU of one or two organisms per ml). Positive cultures represented more than 20 different organisms. Approximately two-thirds of the false-positive results represented mixed flora or pure cultures of less than 5 X 10(4) organisms per ml. The predictive value of a negative result was 98.4%, reflecting a false-negative rate of only 0.7%. No advantages in cost or technician time were noted, but the Lumac method appears to be a useful technique in decreasing reporting time, especially for negative urine cultures.
Accelerating Research
Robert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom
Address
John Eccles HouseRobert Robinson Avenue,
Oxford Science Park, Oxford
OX4 4GP, United Kingdom