Radiation transmission, leakage and beam penumbra measurements of a micro‐multileaf collimator using GafChromic EBT film
Author(s) -
GarcíaGarduño Olivia Amanda,
Celis Miguel Ángel,
LárragaGutiérrez José Manuel,
MorenoJiménez Sergio,
MartínezDávalos Arnulfo,
RodríguezVillafuerte Mercedes
Publication year - 2008
Publication title -
journal of applied clinical medical physics
Language(s) - English
Resource type - Journals
SCImago Journal Rank - 0.83
H-Index - 48
ISSN - 1526-9914
DOI - 10.1120/jacmp.v9i3.2802
Subject(s) - multileaf collimator , collimated light , collimator , optics , linear particle accelerator , materials science , detector , radiosurgery , radiation , leakage (economics) , beam (structure) , physics , nuclear medicine , radiation therapy , laser , medicine , economics , macroeconomics
Micro‐multileaf collimator systems coupled to linear accelerators for radioneurosurgery treatments require a rigorous dosimetric characterization in order to be used in 3D conformal and intensity modulated stereotactic radiosurgery and radiotherapy applications. This characterization involves high precision measurements of leaf transmission, leakage and beam penumbra through the collimation system and requires the use of detectors with high spatial resolution, high sensitivity and practically no energy dependence. In this work the use of GafChromic EBT radiochromic film to measure the basic dosimetric properties of the m3‐mMLC (BrainLAB, Germany) micro‐multileaf collimator system integrated to a 6 MV linear accelerator, is reported. Results show that average values of transmission and leakage radiation are 0.93 ± 0.05 % and 1.08 ± 0.08 % , respectively. The 80–20% beam penumbra were found to be 2.26 ± 0.11 mm along the leaf side (perpendicular to leaf motion) and 2.31 ± 0.11 mm along the leaf end (parallel to leaf motion) using square field sizes ranging from 9.1 to 1.8 cm. These measurements are in agreement with values reported in the literature for the same type of mMLC using different radiation detectors. PACS number: 87.56.N‐